Understanding Emissivity Settings: Why Most Technicians Get Wrong Readings

25.11.26 06:30 AM By Raj Kanabar

Infrared Thermometer Emissivity: Fix Inaccurate Readings


A technician checks a painted bearing housing and then points the same infrared thermometer at a polished section of the adjoining machine. The second reading is surprisingly low. Before reporting that the machine is running cooler, there is a question worth asking: has the temperature changed, or has the surface changed?

An accurate infrared thermometer can still produce an incorrect reading when its emissivity setting does not match the target. This is particularly important in factories where painted motors, stainless steel pipes, copper connections and aluminium components are inspected using the same instrument.

Understanding infrared thermometer emissivity helps technicians distinguish an actual temperature problem from a measurement problem.

What Is Emissivity in an Infrared Thermometer?

Emissivity describes how effectively a surface emits thermal radiation compared with an ideal blackbody at the same temperature. It is expressed as a dimensionless value between 0 and 1.

An ideal blackbody has an emissivity of 1.00. A surface with an emissivity of 0.95 emits 95% of the radiation an ideal blackbody would emit under comparable spectral and viewing conditions. A surface with an emissivity of 0.10 emits only 10% under those conditions.

The thermometer uses its emissivity setting to estimate the temperature from the radiation reaching its detector. It does not automatically identify the material being measured.

If your maintenance team regularly checks different surface finishes, consider infrared thermometers with adjustable emissivity.

Why Does an Infrared Thermometer Show the Wrong Temperature?

For an opaque surface and a short, clear optical path, the detector receives two important contributions: thermal radiation emitted by the target and radiation from the surroundings reflected by that target.

Consider two surfaces at 80°C in a workshop with cooler surroundings. One is painted and the other is polished aluminium. The painted surface generally emits more of its own thermal radiation. The aluminium reflects a greater proportion of the surrounding radiation.

With the same high emissivity setting, the polished aluminium can appear considerably cooler than the painted surface. Put a hot furnace in the reflection path, however, and it may appear hotter than it really is.

A low reading is not always evidence of cooling, and a high reading is not always evidence of overheating.


Is 0.95 the Correct Emissivity Setting for Everything?

No. An emissivity of 0.95 is a useful starting point for many matte, painted and organic surfaces. It is not a universal value for steel, copper or aluminium.

Even two components made from the same alloy can require different settings because of polishing, oxidation, coatings and surface contamination. Temperature, wavelength and viewing angle also influence effective emissivity.

Practical Emissivity Reference Chart

Surface ConditionApproximate Starting Value or BehaviourField Recommendation
Matte industrial paintAround 0.90 to 0.95Verify the coating for critical readings
Black rubberAround 0.95Check the actual material
Dry concreteAround 0.95Check surface moisture and condition
Polished aluminiumCan be near 0.05Prefer a validated reference patch or contact method
Polished copperCan be near 0.05Be especially cautious about reflections
Bare stainless steelHighly finish-dependentEstablish a reference rather than assuming a fixed value
Oxidised or rusted steelOften higher than polished steelVerify the actual oxide layer

These are indicative values, not certified settings for every material. A thermometer whose emissivity adjustment starts at 0.10 cannot directly match a target with an effective emissivity of 0.05. Even instruments that permit lower settings may struggle with highly reflective targets.

How to Set Emissivity Correctly in the Field

Method 1: Use a Known-Emissivity Reference Patch

For a suitable stationary surface, a reference patch is often more practical than guessing a metal's emissivity.

  1. Identify a safe measurement location on the component.

  2. Where permitted, apply a non-reflective reference tape or coating with known emissivity and a suitable temperature rating. Apply it only under safe, isolated conditions.

  3. Allow the patch to reach the temperature of the underlying surface after the equipment returns to a stable operating condition.

  4. Set the IR thermometer to the patch's known emissivity.

  5. Measure the patch from a distance at which it comfortably fills the optical spot.

This gives a reference surface reading, provided the patch does not materially alter the local temperature.

Do not use ordinary adhesive tape on high-temperature equipment, moving parts, exposed live conductors or contamination-sensitive process surfaces.

Method 2: Compare with a Contact Surface Thermometer

Use a suitable calibrated surface probe on a safely accessible, thermally stable component. Allow the reading to stabilise. Measure the same area with the IR thermometer and adjust emissivity until the readings agree, while controlling reflected radiation and keeping the measurement geometry consistent.

Record the resulting setting for that particular surface and operating condition. Agreement at one temperature is not proof that the setting will work across the entire process range.

If reflections cannot be controlled or readings remain unstable, use an appropriate contact method or specialised infrared instrument instead.


Which Infrared Thermometer Should You Buy for Accurate Measurements?

An adjustable emissivity setting is valuable when technicians inspect multiple materials. But the correct purchase also depends on optical resolution, operating environment and temperature range.

ModelEmissivity AdjustmentTemperature RangeD:S RatioBest-Fit Application
UNI-T UT306S0.10 to 1.00-50 to 500°C12:1Routine motor, HVAC and machine inspections
Fluke 62 MAX0.10 to 1.00-30 to 500°C10:1Field maintenance where IP54 protection and 3 m drop-test rating matter
UNI-T UT302D+0.10 to 1.00-32 to 1100°C20:1Higher-temperature targets, smaller measurement areas and recorded inspection readings

For routine checks on painted motor housings, the UT306S may be sufficient. The 62 MAX is worth considering when ruggedness is a priority. The UT302D+ adds a wider temperature range and finer optical resolution at comparable distances.

None of these instruments automatically eliminates errors on shiny metals. Confirm the measurement method before choosing a model based on its stated accuracy or price.


Does a Thermal Imaging Camera Solve the Emissivity Problem?

A thermal camera helps locate temperature variations over a wider area. This is valuable when the technician does not yet know where to point a temperature gun.

For example, the UNI-T UTi260B thermal imager provides 256 × 192 infrared resolution, adjustable emissivity and a measurement range of -20 to 550°C. It can help maintenance teams identify abnormal thermal patterns across motors, machinery and electrical installations.

But a thermal image is still affected by emissivity. A bright area on a polished panel may be a reflection rather than a genuine hot spot. Changing the viewing angle can help reveal a reflection. Where supported, configure reflected apparent temperature as well as emissivity and other relevant measurement parameters.

Compare portable infrared thermal imagers if your maintenance requirement involves locating hot spots rather than checking a known point.

Troubleshooting: When Changing Emissivity Does Not Fix the Reading

SymptomPossible CauseWhat to Check
Small bearing reads cooler than expectedMeasurement spot includes cooler surroundingsTarget size and optical diagram
Reading changes when you move sidewaysReflections or viewing-angle effectsMeasurement geometry and nearby heat sources
Reading drifts after entering a hot plantInstrument has not acclimatisedManufacturer's stabilisation instructions
IR gun disagrees with a vessel probeSurface and internal temperatures differWhether the instruments measure the same quantity
Reading is unstable near a steam lineSteam, dust or condensation in optical pathClear line of sight and clean lens
All reference readings remain incorrectPossible instrument fault or calibration issueControlled reference test and calibration service

At 600 mm, a nominal 12:1 distance-to-spot ratio corresponds to a spot of approximately 50 mm. A 20 mm wide target may therefore be too small.

Consult the actual optical diagram and maintain safe working distances. The laser dot is an aiming aid, not the entire measurement area.


Quick Selection and Inspection Checklist

Before purchasing an accurate infrared thermometer or investigating a measurement error, confirm:

  • Actual surface material, finish and expected temperature range.

  • Required emissivity adjustment range and available reference method.

  • Target dimensions, viewing angle and working distance.

  • Accuracy under relevant conditions, repeatability and response time.

  • Reflected heat sources, ambient temperature and optical obstructions.

  • Enclosure protection, mounting or handheld requirements and display readability.

  • Calibration documentation and any application-specific safety approvals.

  • Whether a spot thermometer, thermal imager or contact instrument is appropriate.

For electrical inspections, only qualified personnel should work under approved electrical safety procedures. Non-contact measurement does not remove electrical or arc-flash hazards.


Final Thoughts

Infrared thermometer emissivity is not a setting to adjust until a reading looks reasonable. It is a measurement parameter that should reflect the surface being inspected.

For repeatable maintenance rounds, document the measurement point, emissivity, distance, viewing angle and machine load. This helps the next technician compare like with like rather than investigate an artificial temperature change.

Need help choosing an infrared thermometer for your application? Contact Radical TechMart with the target material, temperature range and measuring distance for product selection, pricing and availability.

Frequently Asked Questions

1. Why is my infrared thermometer showing a lower temperature than expected?

Incorrect emissivity, reflections of cooler surroundings or an oversized measurement spot can cause low readings. Check the surface finish and compare with a suitable reference before assuming the instrument is faulty.

2. What emissivity setting should I use for an infrared thermometer?

A setting near 0.95 is a useful starting point for many matte surfaces. Use a verified value or reference method for reflective metals and accuracy-critical measurements.

3. What is the correct emissivity of stainless steel?

There is no universal value. Polished, brushed and oxidised stainless steel can behave differently. Determine the effective setting for the actual surface and measurement conditions.

4. Does increasing the emissivity setting increase the temperature reading?

Not necessarily. For a target hotter than its reflected surroundings, increasing the setting generally lowers the calculated temperature for the same received radiation. The relationship can reverse for colder targets.

5. Can an infrared thermometer measure through glass?

Most general-purpose long-wave IR thermometers cannot measure a target reliably through ordinary glass. They mainly detect radiation associated with the glass surface. Use an appropriate infrared window where required.

6. Does a thermal camera need emissivity adjustment?

Yes. Thermal cameras also calculate temperatures from infrared radiation. Correct emissivity and relevant reflection compensation are important for quantitative measurements.

7. Does adjusting emissivity calibrate an infrared thermometer?

No. Emissivity adjustment adapts the calculation to the target. Calibration assesses instrument performance against an appropriate reference radiation source under controlled conditions.

8. Where can I buy an adjustable emissivity infrared thermometer in India?

Radical TechMart offers adjustable-emissivity handheld instruments and thermal imagers. Confirm the model's optical ratio, range, accuracy and suitability for your target before ordering.

Raj Kanabar

Raj Kanabar

Process Automation Strategist RADICAL TECHART SOLUTION PVT. LTD.

Raj Kanabar helps industries measure, monitor, control & improve process parameters.

Items have been added to cart.
One or more items could not be added to cart due to certain restrictions.
Added to cart
Quantity updated
- An error occurred. Please try again later.
Deleted from cart
- Can't delete this product from the cart at the moment. Please try again later.